Application of InSenXe Infrared Moisture Sensor in PVB Film Moisture Testing
2026-06-16
Moisture Content Measurement of Fuyao Hubei Shenghua
Heat Insulation Film (8% Transmittance)
Test three sample pieces with the initial probe setup. Record the displayed values (as shown in the left column) and input the corresponding sample values (as shown in the right column).

- Calibration Results
The correlation coefficient is 0.9995 and the standard deviation is 0.0181, demonstrating highly accurate test results.
- Post-calibration Measurements
| Sample Values | Displayed Values |
| 0.28 | 0.25 |
| 0.38 | 0.39 |
| 0.61 | 0.61 |
Kingboard Transparent Film Moisture Content Testing
- Pre-calibration Data Collection and Input

- Calibration Results

- Post-calibration Measurements
| Sample Values | Displayed Values |
| 0.24 | 0.24 |
| 0.39 | 0.38 |
| 0.60 | 0.60 |