Application of InSenXe Infrared Moisture Sensor in PVB Film Moisture Testing

2026-06-16

Moisture Content Measurement of Fuyao Hubei Shenghua

 Heat Insulation Film (8% Transmittance) 

Test three sample pieces with the initial probe setup. Record the displayed values (as shown in the left column) and input the corresponding sample values (as shown in the right column).

  • Calibration Results

 

The correlation coefficient is 0.9995 and the standard deviation is 0.0181, demonstrating highly accurate test results. 

  • Post-calibration Measurements 
Sample ValuesDisplayed Values
0.280.25
0.380.39
0.610.61

Kingboard Transparent Film Moisture Content Testing 

  • Pre-calibration Data Collection and Input 

  • Calibration Results

  • Post-calibration Measurements 
Sample ValuesDisplayed Values
0.240.24
0.390.38
0.600.60